Isolated Process Control Signal Conditioning Products
SCM7B
For information call 800-444-7644
84
SCM7B47
Isolated Linearized Thermocouple Input Modules
Description
SCM7B47 modules accept a single channel of input from Type J, K, T, E, R,
S, B, or N thermocouples. The signal is filtered, isolated, amplified, linearized,
and converted to a high level analog voltage for output to the process control
system (Figure 1).
Linearization is achieved by creating a non-linear transfer function through the
module itself; refer to AN505. This non-linear transfer function is configured at
the factory and is designed to be equal and opposite to the thermocouple
non-linearity.
Cold junction compensation (CJC) is performed using an NTC thermistor (see
Additional SCM7B Part Numbers section for P/N and AN701 for further
information) externally mounted under the field-side terminal block on the
backpanel (Figure 1). Open thermocouple detection is upscale using a 30nA
current source in the input circuitry.
These modules incorporate a five-pole filtering approach to maximize both time
and frequency response by taking advantage of both Thomson (Bessel) and
Butterworth characteristics. One pole of the filter is on the field side of the
isolation barrier; four are on the process control system side.
After the initial field-side filtering, the input signal is chopped by a proprietary
chopper circuit and transferred across the transformer isolation barrier,
suppressing transmission of common mode spikes and surges. The signal is
then reconstructed and filtered for process control system output.
Modules accept a wide 14 - 35VDC power supply range (+24VDC nominal).
Their compact packages (2.13"x1.705"x0.605" max) save space and are ideal
for high channel density applications. They are designed for easy DIN rail
mounting using any of the -DIN backpanels.
Figure 1: SCM7B47 Block Diagram
Features
 Interfaces to Type J, K, T, E, R, S, B and N
Thermocouples
 Linearizes Thermocouple Signals
 Provides High Level Voltage Outputs
 1500Vrms Transformer Isolation
 Accuracy, ±0.06% to ±0.16% of Span Typical
 ANSI/IEEE C37.90.1 Transient Protection
 Input Protected to 120Vrms Continuous
 Noise, 1mV Peak (5MHz), 500µV RMS
(100kHz)
 CMRR, 160dB
 NMR, up to 85dB
 Easy DIN Rail Mounting
 CSA Certified, FM Approved
 CE and ATEX Compliant
Refer to AN701 for V
out
to
temperature conversion information
Vista de pagina 0
1 2

Indice de contenidos

Pagina 1 - Features

Isolated Process Control Signal Conditioning ProductsSCM7BFor information call 800-444-764484SCM7B47Isolated Linearized Thermocouple Input ModulesDesc

Pagina 2 - Ordering Information

SCM7BVisit our website www.dataforth.com85Isolated Process Control Signal Conditioning ProductsSCM7BSpecificationsInputSignal Range Thermocouple(1)(Se

Comentarios a estos manuales

Sin comentarios